SCICAN STATIM消毒器密封圈安装和保养说明(英文)SCF006SCS001
Installation and monthly maintenance for 安装和每月的维护
The rpi cassette seal (rpi part #scs001) to fit the statim 2000cassette autoclave 2000高压蒸汽灭菌器
The rpi cassette seal (rpi part #scs029) to fit the statim 5000cassette autoclave 5000高压蒸汽灭菌器
1. The cassette must be opened and cooled before removing the cassette seal.(See Figure A)1。
2. Start with any of the(4) corners and begin prying the cassette seal loose using the RPI Cassette Seal Removal Tool (RPI Part #RPT372) and by pulling down and out at the same time.(See Figures B and C)
3. Once the first corner is free, take hold of the cassette seal and pull down and out in order to remove it from the cassette groove. Be careful not to tear the (2) locating tabs in each corner of the cassette seal.(See Figures B and C)
4. Once the cassette seal has been removed, it is important to inspect, clean and remove any residue and /or debris from the cassette groove. Check for dings and dents on both halves of the cassette; if present, a proper seal may not occur.(See Figure D)
5. Using(1) package of RPI Liquid Soap (RPI Part #RPS287),lube the cassette seal completely. Note: It is not necessary to lube the Inside of the cassette seal.(See Figure E)
6. Now you are ready to install the casette seal
7. When installing the cassette seal,it is best to start in the back corner ,nearest the holes of cassette.
Slip the cassette seal under into the corner ,placing the locting tabs at the edge of the open corner.Now align the holes and their locating tabs with the cassette,The locating tabs must proture throng the holea of the casstte (see figure f)
8 insallthe other corner in the same manner .now insert the cassette seal into the groove of the cassette -one side at a time (see figures g ahd h )
9As you move along onstalling the cassette seal ,check the corners to ensure that the locating tabs stay in place.Note:the sidea should be inserted until smooth-without waves or bumps(see figure)
10 After you have completely installed the cassette seal ,make a final inspecting of all corners soap from the holes as the soap can cause a build up on the probes and in tuin cause a bad seal
11 Run a few test cycles to check fou steam leaks .Note:during the test cycles ,it is normal to hear a hissing soud at first.The soud is from the excess liquid soap expelling and the cassette seal seating in place.
12 After running the test cycles ,if steam still leaks ,reinstall the cassette seal ahd tun addtional cycles .If steam leaks still occur ,it is possible that the problem was not a faulty cassette seal ,but something else,so refer to the operator s manusl or contact your service professional.
Usage tips:使用技巧
1.Always note the position of the cassette when inserting it into unit-if the cassette is installed upsidedown,the thermocouple will bend and the sterilizer will become inoperative.
2.Lubricate the exposed area of the cassette seal with RPI Liquid Soap (RPI Part # RRS287) on a weekly basis
3.Alway keep the inside of the casstte clean.Remove residue and/or debris from the metal edge of the lower half of the cassette on a regular basis.
4.Watch for steam leaks from the cassette as they can cause damage to the microswith ,thus rendering the sterilizer inoprative.
5.Refill the reservoir with distilled water only.At the same time that the reservior is refilled.the condenser waste bottle should be emptied and refilled.Always refill the condenser waste bottle with tap water to the Minimum Fill Level as shown on the bottle(Note:Never operate sterilizer when water level is above the Maximum Fill Line or below the Minnum Fill Line)
2000 Replacement Parts Industries ,Inc,SCS001 INS REV B (08/04)
环氧乙烷指示剂 型号1264指示剂
美国3M环氧乙烷指示剂
由枯草杆菌黑色变种芽孢,培养基及塑料外壳整合组成。

上海爱宝医疗,AIBAOYL,型号全,价格低。
公司地址:上海市长逸路15号复旦软件园B幢812室
电话:13681661622 021-66187083 021-66187055 021-66187008 021-56149570
传真:021-66145903
联系人:王经理 15026564022 021-66187083
电子邮件:15026564022@163.com 1435024644@qq.com
QQ:小王 1435024644
爱宝医疗网:http://www.aibaoyl.cn
爱宝医疗网:http://www.aibaoyl.cn